      Tamoxifen Preparation and Treatment in Mice for Genetic Recombinations

      0

      Tamoxifen Preparation and Treatment in Mice for Genetic Recombinations

      Rationale: Tamoxifen is used to kick-start genetic recombinations in genetically engineered mouse models (GEMMs) and to activate genes through a cascade of events.

      Here is a detailed step-by-step guideline on how to prepare and inject Tamoxifen into mice. Please ensure you are properly trained and authorized to handle laboratory mice before proceeding. Always follow animal welfare laws and your facility’s study protocol.

      Tamoxifen Preparation

      1. Calculate the appropriate concentration of tamoxifen for your experiment; the dose usually falls in the range of 40–100 mg/kg/day, but adjust according to the specifics of your study.

      2. Tamoxifen is soluble in ethanol at 50 mg/mL. To create this stock solution, dissolve the required amount of tamoxifen in 100% ethanol.

      3. After completely dissolved, dilute the stock solution with corn or sunflower oil to reach the intended final concentration.

      4. Store it at -20 degrees Celsius for long-term storage.

      Intraperitoneal Injection

      1. Allow the tamoxifen solution to reach room temperature before the injection or warm at 370C (recommended) before use to match the internal body temperature of mice for improved function.

      2. Load a sterile insulin syringe (1 mL) with the appropriate volume to achieve the desired dosage.

      3. For intraperitoneal (IP) injection, the mouse is restrained manually or using a suitable restrainer. Applying gentle pressure on both sides of the lower part of the abdomen makes the injection easier and more accurate.

      4. Hold the syringe and needle like a dart and puncture the skin and peritoneum on the lower right side of the abdomen, avoiding vital organs.

      5. Once inserted, depress the syringe slowly to inject the solution. Do not force it if you meet resistance.

      6. Properly manage the needle after injection to prevent injuries and infections.

      7. Monitor mice after the injection and record any changes in their behavior, weight, or appearance.

      Be aware that tamoxifen administration can induce some side effects in mice, such as weight loss or pyometra in females; monitor them regularly.

      Usage Example:

      At 6 weeks of age, mice were intraperitoneally injected with 75 mg per kg (body weight) of tamoxifen (Sigma-Aldrich, T5648-5G) dissolved in corn oil once a day for 5 consecutive days (Yin et al., 2023, Nat. Immunol.).

      This procedure must be done accurately and carefully. Future adjustments must be made based on the observed results. Always follow your Institutional Animal Care and Use Committee regulations and guidelines. If any issues occur, consult with your attending veterinarian.

      Remember, this guide does not replace professional training, and an experienced investigator or veterinarian should supervise all steps until proficiency is achieved.

      DNA Quantification using Qubit Fluorometer

      1

      DNA Quantification using Qubit Fluorometer

      The Qubit Fluorometer is a device from Thermo Fisher Scientific used to quantify DNA, RNA, and protein in samples by using a fluorescent dye that binds to these molecules. The Qubit uses specialized assay kits, like the Qubit dsDNA HS (High Sensitivity) Assay Kit, for quantifying double-stranded DNA (dsDNA).

      Here is a step-by-step process of using the Qubit fluorometer for DNA quantification:

      Materials Required: Qubit fluorometer, Qubit dsDNA HS assay kit (or similar), Qubit assay tubes, Pipettes and tips, and DNA samples.

      1. Prepare Qubit working solution: The working solution is made by diluting the Qubit dsDNA HS Reagent in the Qubit dsDNA HS Buffer. The standard dilution is 1 part reagent to 199 parts buffer. Mix the reagent before use as it may be prone to forming precipitates.

      2. Prepare DNA Standards: The Qubit assay kit includes two DNA standards (Standard #1 and #2). These standards are used to calibrate the fluorometer. Add 190 µL of the working solution to each standard tube and then add 10 µL of the respective standard. Mix gently, then read the standards on the Qubit fluorometer.

      3. Prepare DNA samples: Add 198 µL of the working solution into an assay tube for each DNA sample. Then, add 2 µL of the DNA sample. Mix the solution gently.

      4. Measure DNA Standards and Samples: Place the tube containing Standard #1 into the Qubit and close the lid. On the screen of the Qubit, select “DNA” then “High Sensitivity” (or the appropriate assay) and tap “Read”. The fluorometer will read the standard and then prompt you to insert Standard #2. Then, proceed to read your samples.

      5. Interpret the results: Once your sample is read, the Qubit will display the DNA concentration in your sample on the screen. The Qubit fluorometer software also allows you to export your data for further analysis.

      Remember to follow good lab practices to prevent contamination of your samples. In addition, to maintain the stability of the fluorescent dye reagent, avoid extended periods of light exposure.

      DNA Quantification using Nanodrop spectrophotometer

      1

      DNA Quantification using Nanodrop spectrophotometer

      The nanodrop spectrophotometer is commonly used to quantify DNA concentration and assess purity. Here is a basic outline of the process:

      1. Calibrate the machine: It’s usually done by using a blank solution that the machine will use as a base reading. This is typically the same solution your DNA is suspended in, such as pure distilled water or TE buffer.

      2. Measure the Sample: Once the Nanodrop is calibrated, you can measure your sample. Apply a small amount (1-2 µl) of your sample to the measurement pedestal.

      3. Take the Reading: Close the arm of the Nanodrop and take the reading. The machine will provide you with measurements at various wavelengths that can be used to quantify and qualify your sample.

      The key measurements you’ll find in the Nanodrop report are:

      – 260 nm: This is the absorbance value of your DNA. The concentration of your DNA (ng/µl) is calculated based on this value.

      – 260/280 ratio: This ratio gives you an indication of the purity of your DNA. Pure DNA generally has a ratio of around 1.8, while lower ratios may indicate protein contamination.

      – 260/230 ratio: This is another measure of purity. The optimal value tends to be above 2.0. A lower ratio could indicate organics, salts, or other compounds contamination.

      Remember, while Nanodrop is a fast and easy method for quantifying DNA, it might not be the most accurate when it comes to very low concentrations or when the sample is contaminated. In such cases, fluorescence-based quantification methods like Qubit might be more appropriate.

      Privacy Overview

      This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.